The flow rate programmed on the console may be changed without automatically updating the flow rate programmed on the power head. The injector will always inject in accordance with the parameters shown on the power head. If this lack of synchronization occurs and is not noticed by the clinican, contrast would be injected at a higher or lower flow rate than desired.
Source facts
Action
The recalling firm sent a letter dated 11/7/2006 to all consignees.
Initiated
2006-11-07
Root cause
Other
Terminated
2007-07-21
Distribution
Worldwide distribution ---- including USA and countries of Australia, Belgium, Brazil, Canada, Chile, Columbia, Denmark, Mexico, Panama, Puerto Rico, Singapore and South Africa.
Recall number
Z-0288-2007
Recall status
Terminated
CDRH record ID
49391
FDA classified
2006-12-16
Recall event number
36811