Skip to content

Public FDA record

Z-0288-2024: EVIS EXERA III Gastrointestinal Videoscope

Recall record Z-0288-2024 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0288-2024
Company in FDA record
Olympus Corporation of the Americas
FDA classification date
Nov 9, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Sterilization failures when devices are sterilized per the Reprocessing Manual due to lack of drying time for the endoscope channel.

Source facts

Action
The firm distributed recall notifications via mail to consignees dated 10/02/2023. Consignees are asked to inspect inventory for affected devices, ensure that all reprocessing personnel are aware and knowledgeable of the recall notice and its contents, and acknowledge receipt of the recall notification by returning the enclosed response form via email to olympus5381@sedgwick.com or by fax at 800-913-4345. Customers with any questions about the acknowledgement form are to call 866-912-9544. Customers with any questions about the recall are to contact Cynthia Ow by phone at 647-999-3203 or by email at Cynthia.Ow@Olympus.com.
Initiated
2023-10-02
Root cause
Reprocessing Controls
Distribution
Nationwide Distribution.
Recall number
Z-0288-2024
Recall status
Open, Classified
CDRH record ID
203663
FDA classified
2023-11-09
Recall event number
93212

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗