Open original FDA record ↗Recall record
- FDA identifier
- Z-0288-2024
- Company in FDA record
- Olympus Corporation of the Americas
- FDA classification date
- Nov 9, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Sterilization failures when devices are sterilized per the Reprocessing Manual due to lack of drying time for the endoscope channel.
Source facts
- Action
- The firm distributed recall notifications via mail to consignees dated 10/02/2023. Consignees are asked to inspect inventory for affected devices, ensure that all reprocessing personnel are aware and knowledgeable of the recall notice and its contents, and acknowledge receipt of the recall notification by returning the enclosed response form via email to olympus5381@sedgwick.com or by fax at 800-913-4345. Customers with any questions about the acknowledgement form are to call 866-912-9544. Customers with any questions about the recall are to contact Cynthia Ow by phone at 647-999-3203 or by email at Cynthia.Ow@Olympus.com.
- Initiated
- 2023-10-02
- Root cause
- Reprocessing Controls
- Distribution
- Nationwide Distribution.
- Recall number
- Z-0288-2024
- Recall status
- Open, Classified
- CDRH record ID
- 203663
- FDA classified
- 2023-11-09
- Recall event number
- 93212