Open original FDA record ↗Recall record
- FDA identifier
- Z-0289-2008
- Company in FDA record
- Metrika Inc
- FDA record creation date
- Jun 25, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Incorrect Test Results: Due to manufacturing errors regarding temperature compensation factors, the device will give erroneously high results if the test is performed below 70 degrees Fahrenheit.
Source facts
- Action
- A September 14, 2007, Urgent Device Recall letter was issued to all Distribution Center Managers to return recalled product for replacement. Customer notification was made via a Dear Healthcare Professional letter dated Sept. 14, 2007, also titled Urgent Medical Device Recall.
- Initiated
- 2007-08-31
- Root cause
- Process control
- Terminated
- 2008-03-25
- Distribution
- Nationwide including states of AK, AL, CA, CO, CT, DC, FL, GA, HI, IL, IN, KS, KY, MA, MD, MI, MN, MS, NC, NJ, NV, NY, OH, OK, PA, SC, TN, TX, UT, VA, WA, WI and Puerto Rico.
- Recall number
- Z-0289-2008
- Recall status
- Terminated
- CDRH record ID
- 64351
- FDA record created
- 2015-06-25
- Recall event number
- 44752