Skip to content

Public FDA record

Z-0289-2017: Vari-Safe Injection Needle US endoscopy 5976 Heisley Road Mentor, Ohio 44060 A sterile, single use, flex

Recall record Z-0289-2017 for US Endoscopy Group Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0289-2017
Company in FDA record
US Endoscopy Group Inc
FDA classification date
Oct 19, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

US Endoscopy is conducting a voluntary product recall of one lot of the Vari-Safe Injection Needle as it was identified that the lot was incorrectly distributed. The lot was assembled with internal catheter components which differ from the original Vari-Safe needle internal catheter components. All components are dimensionally similar to the original Vari-Safe needle internal catheter components and the components have history of use on other US Endoscopy needle products.

Source facts

Action
US Endoscopy sent an Urgent- (Vari Safe) injection needle recall letter dated July 6, 2016. US Endoscopy requests that you: 1.Quarantine the affected Vari-Safe Injection devices in your stock. 2. Destroy each affected device by opening the device package, cutting the device catheter with a scissors, and then discarding the device and package. 3.Complete the attached Customer Response Card and return to the attention of US Endoscopy Customer Service, via fax 1-440-639-4495 or email orders@us endos copy.com . A no charge replacement will be provided for each unused Vari-Safe Injection Needle device from the affected lot. We apologize for any inconvenience this matter may cause you, and as always, US Endoscopy is dedicated to supporting our products and valued Customers. For further questions please call (800) 769-8226.
Initiated
2016-05-18
Root cause
Other
Terminated
2017-05-10
Distribution
US Distribution to the states of : OH, OR, WA and Internationally to the country of Italy.
Recall number
Z-0289-2017
Recall status
Terminated
CDRH record ID
147998
FDA classified
2016-10-19
Recall event number
74590

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗