Z-0290-2007: Medtronic Kinetra Dual Program Neurostimulator for Deep Brain Stimulation, model 7428. Rx only. Medtron

Recall record Z-0290-2007 for Medtronic Neurological. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0290-2007
Company in FDA record
Medtronic Neurological
FDA classification date
Dec 19, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

A subset of Kinetra implantable neurostimulators may experience a failure of wire connections between the electronic hybrid circuit and battery which may lead to sudden cessation of therapy. Sudden cessation of therapy can result in the immediate return or worsening of underlying symptoms due to the progression of the disease state.

Source facts

Action
Letter dated July 2006. Recommendations were made to Health Care Providers of the following: 1.Consider the potential for this anomaly if your patient present with a device in a POR state. 2. Consider informing your patients of this risk and advise them to seek attention immediately if they experience a return of pre-implant or more advanced symptoms. 3. Determine whether device replacement is warranted based upon consultation with your patient, review of the patient''s medical history and consideration of the relative risks of abrupt cessation of stimulation versus the risk of a device replacement surgical procedure. 4. Report any malfunction, removal and replacement to Medtronic and FDA suing the medical device reporting system. 5. Acknowledge this communication using the reply form enclosed.
Initiated
2006-07-19
Root cause
Other
Terminated
2007-04-18
Distribution
Worldwide, including USA, Puerto Rico, Australia, Austria, Belgium, Canada, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Israel, Italy, Jordan, Lebanon, Luxembourg, Netherlands, Norway, Poland, Portugal, Russia, Saudi Arabia, Slovakia, South Africa, Spain, Sweden, Switzerland, Turkey, United Kingdom.
Recall number
Z-0290-2007
Recall status
Terminated
CDRH record ID
47276
FDA classified
2006-12-19
Recall event number
35967

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗