Open original FDA record ↗Recall record
- FDA identifier
- Z-0290-2007
- Company in FDA record
- Medtronic Neurological
- FDA classification date
- Dec 19, 2006
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
A subset of Kinetra implantable neurostimulators may experience a failure of wire connections between the electronic hybrid circuit and battery which may lead to sudden cessation of therapy. Sudden cessation of therapy can result in the immediate return or worsening of underlying symptoms due to the progression of the disease state.
Source facts
- Action
- Letter dated July 2006. Recommendations were made to Health Care Providers of the following: 1.Consider the potential for this anomaly if your patient present with a device in a POR state. 2. Consider informing your patients of this risk and advise them to seek attention immediately if they experience a return of pre-implant or more advanced symptoms. 3. Determine whether device replacement is warranted based upon consultation with your patient, review of the patient''s medical history and consideration of the relative risks of abrupt cessation of stimulation versus the risk of a device replacement surgical procedure. 4. Report any malfunction, removal and replacement to Medtronic and FDA suing the medical device reporting system. 5. Acknowledge this communication using the reply form enclosed.
- Initiated
- 2006-07-19
- Root cause
- Other
- Terminated
- 2007-04-18
- Distribution
- Worldwide, including USA, Puerto Rico, Australia, Austria, Belgium, Canada, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, India, Israel, Italy, Jordan, Lebanon, Luxembourg, Netherlands, Norway, Poland, Portugal, Russia, Saudi Arabia, Slovakia, South Africa, Spain, Sweden, Switzerland, Turkey, United Kingdom.
- Recall number
- Z-0290-2007
- Recall status
- Terminated
- CDRH record ID
- 47276
- FDA classified
- 2006-12-19
- Recall event number
- 35967