Skip to content

Public FDA record

Z-0290-2017: RTI Surgical Streamline OCT Occipito-Cervico-Thoracic Spinal Rod, CoCr, 3.2mm (D) x 240mm (L) Product Us

Recall record Z-0290-2017 for RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.). View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0290-2017
Company in FDA record
RTI Surgical, Inc. (dba Pioneer Surgical Technology, Inc.)
FDA classification date
Oct 20, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

RTI Surgical Inc. (RTI) is conducting a voluntary recall on two specific lots of the Streamline OCT Straight Rod, CoCr. 3.2mm (D) x 240mm (L) due to the potential for small cracks to be present on the rods. The small cracks on the rods used in implants could cause reduced strength of these affected implants, and lead to patient injury.

Source facts

Action
RTI issued e-mail cutsomer notices for their voluntary recall of lots 234829 and 234829 of the Streamline OCT Straight Rod, CoCr, 3.2mm (D) x 240mm (L) on 09/16/2016. RTI Surgical has produced replacement implants that have been verified as not containing any cracks. These implants will be sent to customers once they have identified how many replacements are needed. Please follow the below instructions. 1. Review inventory located in your facility and any other point of use locations that you may be responsible for. Immediately quarantine any implants listed in this recall notice. Please refer to the pictures included in customer notices for the etch location of the lot numbers for these implants: 2. Contact the RTI Surgical Field Inventory representative Jennifer Mannor so that replacements can be sent and an RGA and return shipping labels can be created if needed. 3. Review, complete and return the attached recall reply form to Dan Nelson (dnelson@rtix.com) in accordance with the directions on the business reply form within 5 business days of receipt of this notification. Should customers have any questions please do not hesitate to contact RTI Surgical Field Inventory representative Jennifer Mannor (Email: jmannor@rtix.com Phone: 906-225-5872.
Initiated
2016-09-16
Root cause
Process control
Terminated
2017-06-01
Distribution
US Nationwide Distribution in the states of: AL, CA, CO, FL, IN, IL, MI, MO, MS, NC, NY, OK, TN, TX, and VA,.
Recall number
Z-0290-2017
Recall status
Terminated
CDRH record ID
149750
FDA classified
2016-10-20
Recall event number
75242

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗