Z-0290-2024: Cardiohelp-i, blood oxygenation and carbon dioxide removal system, Part Numbers 701072780 (Cardiohelp-i),

Recall record Z-0290-2024 for Maquet Medical Systems USA. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0290-2024
Company in FDA record
Maquet Medical Systems USA
FDA classification date
Nov 9, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It was found that the fixation (weld) employed by the supplier of the nut bolt to the slide rail plate was insufficient. The deficient weld allows the nut bolt to break off from the slide rail plate with slight manual force.

Source facts

Action
An URGENT MEDICAL DEVICE Correction notification letter dated 10/12/23 was sent to customers. Actions to be taken by the Customer: Our records indicate that you have received a CARDIOHELP-i System having a serial number that is affected by this voluntary Medical Device Correction. " Please examine your inventory immediately to determine if you have any affected CARDIOHELP System(s) in your inventory. " Should you have any affected CARDIOHELP-i System(s), your unit(s) will need to be returned to Getinge for correction. Getinge will contact your facility to arrange the return of your unit for correction. A CARDIOHELP loaner can be provided if needed. " Please do not to fixate any accessory at the slide rail until the slide rail holders have been replaced. " Affected CARDIOHELP-i can be used for intra-hospital transport only when placed on a Sprinter cart. " Inter-hospital transport is not permitted until the slide rail holders have been replaced. " Please forward this information to all current and potential CARDIOHELP-i users within your hospital / facility. " If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action. " Whether or not your facility has affected product(s) listed in this notice, please complete and sign the attached MEDICAL DEVICE- CORRECTION RESPONSE FORM (Page 4) to acknowledge that you have received this notification. Return the completed form to Getinge by e-mailing a scanned copy to Cardiohelpsliderail2023.act@getinge.com or by faxing the form to 1-866-313-4049. Type of Action by Getinge: A Getinge representative will contact you to arrange for the repair of your CARDIOHELP -i unit(s). Getinge will correct your CARDIOHELP -i unit(s) upon receipt and will return your unit(s) once correction has been completed. This voluntary correction only affects the products listed on Page 1; no other products are affected by this voluntary correction. We apologize for
Initiated
2023-10-12
Root cause
Device Design
Distribution
US Nationwide.
Recall number
Z-0290-2024
Recall status
Open, Classified
CDRH record ID
203278
FDA classified
2023-11-09
Recall event number
93013

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗