Z-0291-2021: GORE TAG Conformable Thoracic Stent Graft with ACTIVE CONTROL System

Recall record Z-0291-2021 for W L Gore & Associates, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0291-2021
Company in FDA record
W L Gore & Associates, Inc.
FDA record creation date
Oct 23, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When the delivery catheter was attempted to be removed, there was an observed connection between the stent graft and the delivery catheter that prevented the delivery catheter from being withdrawn.

Source facts

Action
On 09/23/2020, the firm sent an "URGENT Medical Device Safety Correction" via First Class Postal (USPS) informing them that all lots of its Thoracic Stent Graft Systems may encounter difficulty withdrawing the delivery catheter due to a proximal stent apex being lodged with the delivery catheter leading olive. Customers are instructed to: Immediate Actions for the Recipient: - Take note of amendment/reinforcement of Instructions For Use (IFU) - Please respond to the enclosed acknowledgement - Please share this letter with others in your hospital or clinic as appropriate The Recalling Firm is providing physicians with this information so that appropriate risk-related decisions can be made with the patient when considering the Conformable Thoracic Stent Graft System. The Recalling Firm maintains its confidence in the safety and efficacy of the device and will not be removing the affected products from the market. There are no actions required for patients already implanted with the Thoracic Stent Graft. Please contact the Recalling Firm's Customer Service (email: MPDCustomerCare@wlgore.com or by phone at 800-528-8763) with any questions related to this letter.
Initiated
2020-09-09
Root cause
Process control
Distribution
U.S. - Nationwide O.U.S. - Australia, Austria, Belgium, Croatia, Cypress, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Ireland, Israel, Italy, Japan, Lebanon, Luxembourg, Monaco, Netherlands, Norway, New Zealand, Poland, Portugal, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, and United Kingdom
Recall number
Z-0291-2021
Recall status
Open, Classified
CDRH record ID
183678
FDA record created
2020-10-23
Recall event number
86421

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗