Skip to content

Public FDA record

Z-0291-2027: ALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000V

Recall record Z-0291-2027 for Canon Medical System, USA, INC.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0291-2027
Company in FDA record
Canon Medical System, USA, INC.
FDA classification date
Oct 27, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to software issue, X-ray interventional system may experience internal error if F-Store or imagine acquisition is performed while clinical analysis function is use.

Source facts

Action
On October 12, 2026, Canon Medical Systems USA, Inc. issued a Urgent: Medical Device Correction to affected consignees. Canon provided the following to consignees: 1. A representative of Canon Medical Systems USA will contact you to schedule a time to conduct installation of modified software to prevent internal errors from occuring if F-STORE or image acquisition is performed while CAAS clinical analysis function is used. 2. If you have any questions regarding this matter, please contact your service representative or contact our InTouch team at (800) 521-1968 or by email intouchdispatch@us.medical.canon. 3. In order to comply with the FDA record keeping requirement, Canon respectfully requests that the attached form be completed and returned via email to fieldcorrections@us.medical.canon
Initiated
2026-10-12
Root cause
Software design
Distribution
US Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, HI, IL, LA, MA, MI, MN, MS, NC, NV, OH, OR, TX.
Recall number
Z-0291-2027
Recall status
Open, Classified
CDRH record ID
222830
FDA classified
2026-10-27
Recall event number
99889

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗