Skip to content

Public FDA record

Z-0292-2021: 9372HD Digital Capture (ver. 1.1.0 + w/ 9263 endoPortal),

Recall record Z-0292-2021 for Pentax of America Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0292-2021
Company in FDA record
Pentax of America Inc
FDA classification date
Oct 23, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is an intermittent software issue that could affect the systems, in which an exam video for one patient (Patient A) might be copied to another patient (Patient B). This does not occur at the time the user performs and initially reviews an exam; it is not evident until a follow-up review occurs.

Source facts

Action
URGENT MEDICAL DEVICE CORRECTION notification letters were sent on 9/22/20. Customer Instructions: Enclosed with this letter is a Field Correction Response Form which identifies the serial numbers of the affected products your facility owns. PENTAX Medical will be taking a two step approach to contain and correct the problem. o First, we will contact you to schedule a time for our Field Service to temporarily disable file transfer to endoPortal. o During the time that file transfers to endoPortal are disabled, you can continue to create and save exam video files. When this issue is resolved, all files recorded during that period will be uploaded and will be available for endoPortal. o Second, PENTAX Medical is developing a software solution that will be deployed remotely. We will coordinate all updates with your IT administrator. Adverse events experienced with the use of this product should be reported as soon as possible to PENTAX at vigilance@pentaxmedical.com. Contact Information: PENTAX regrets any inconvenience that this action may cause and appreciates your understanding and cooperation. PENTAX will issue additional communications as further information becomes available. Please be assured that maintaining patient safety and quality is our utmost priority. If you have any questions regarding this action, please feel free to contact us at: Tel: 1-800-431-5880 (8:30 AM 5:00 PM, Monday Friday, EST) Fax: 201-799-4063 (alternate 201-391-4189) Email: customeradvisories@pentaxmedical.com. An Urgent Medical Device Correction notification letter dated 10/22/20 was sent to customers. This letter is a clarification of and replaces the letter dated September 2, 2020, which you received previously. This letter is to inform you that PENTAX Medical (PENTAX) is conducting a voluntary corrective action of impacted 9372HD Digital Capture version 1.1.0 or higher and 9310HD Digital Video Capture Module with software version 3.4.0 or
Initiated
2020-09-22
Root cause
Software design
Terminated
2022-05-16
Distribution
US Nationwide and Canada
Recall number
Z-0292-2021
Recall status
Terminated
CDRH record ID
183712
FDA classified
2020-10-23
Recall event number
86440

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗