Skip to content

Public FDA record

Z-0293-04: IMx Folate Reagent Pack, list 2220-88 (100 tests - reagents packaged with Ion Capture Reaction Cells) and

Recall record Z-0293-04 for Abbott Laboratories HPD/ADD/GPRD. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0293-04
Company in FDA record
Abbott Laboratories HPD/ADD/GPRD
FDA classification date
Jan 6, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Chemotherapeutic concentrations of methotrexate in patient samples may carryover to subsequent samples, which would result in falsely elevated folate values, when performing IMx Folate testing.

Source facts

Action
Abbott notified their international affiliates via e-mail on 11/23/03. Important Product Information letters dated 11/24/03 will be sent to the end users by the affiliates, informing them of the possible methotrexate carryover to subsequent samples and falsely elevated folate values.
Initiated
2003-11-23
Root cause
Other
Terminated
2004-07-22
Distribution
There was no U.S. distribution. The product was internationally distributed through Abbott subsidiaries in Colombia, Chile, Uruguay, Argentina, Brazil, Canada, England and Australia.
Recall number
Z-0293-04
Recall status
Terminated
CDRH record ID
30609
FDA classified
2004-01-06
Recall event number
27835

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗