A component in the Certitude Loader may have a
molding overflow defect that could detach during advancement of the delivery system and
potentially embolize from the loader into the patient.
Source facts
Action
Recall notification letters were sent to affected consignees via Fed-Ex on 7/21/2017.
Initiated
2017-07-21
Root cause
Employee error
Terminated
2019-09-13
Distribution
United States: All states with the exception of: Alaska, Hawaii, Idaho, New Mexico, Rhode Island, South Dakota and Wyoming
Recall number
Z-0293-2018
Recall status
Terminated
CDRH record ID
160257
FDA classified
2018-01-10
Recall event number
78738