Skip to content

Public FDA record

Z-0293-2022: Medtronic Wireless Recharger Model WR9220 included in the RS5200 Recharger Kit used by the InterStim Micr

Recall record Z-0293-2022 for Medtronic Neuromodulation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0293-2022
Company in FDA record
Medtronic Neuromodulation
FDA record creation date
Nov 24, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The wireless rechargers (WR; WR9200 and WR9220; included in the Recharger Kits) become unresponsive if the charging steps, indicated in the Recharger Patient User Guide, are not followed.

Source facts

Action
The recalling firm issued undated letters on 11/4/2021 via USPS Priority Shipping. The recalling firm has three different letters addressed to the Healthcare Professional depending upon the type of device the wireless recharger is used with (InterStim Micro, product number RS5200; Activa RC, product number RS6200; or RestoreUltra & RestoreSensor, Product number RS7200, which is currently not distributed in the U.S. at this time). The letter gives an issue overview and provides requested actions. Those actions include reinforcing the instructions in the enclosed Patient Quick User Guide with their patients and to attempt to reset the WR if the device appears unresponsive. If the problem is not resolved, then contact information is provided for Medtronic Technical Services. A customer confirmation form dated November 2021 was enclosed to acknowledge the consignee has received the letter.
Initiated
2021-11-04
Root cause
Other
Distribution
US Nationwide, including PR. There was government distribution but no military distribution was identified.
Recall number
Z-0293-2022
Recall status
Open, Classified
CDRH record ID
190094
FDA record created
2021-11-24
Recall event number
88953

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗