Skip to content

Public FDA record

Z-0294-05: Universal Cable Adaptor (Part #17765-001and labeled on outer package as part no. 17820-001) distributed f

Recall record Z-0294-05 for Pulmonetic Systems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0294-05
Company in FDA record
Pulmonetic Systems, Inc.
FDA classification date
Dec 8, 2004
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Reports have been received indicating that the Universal Cable Adaptor, intended to correct an earlier Class I recall of LTV Series Ventilators due to their inability to continue operating when external power source is switched to internal battery, is not functioning as intended and/or the cable is not securely attaching to the connection on the ventilator.

Source facts

Action
Beginning November 19, 2004 consignees who received the Universal Cable Adaptors (UCA) as correction to the voluntary November 1, 2004 Class I Recall were notified by telephone to discontinue installation of adaptors in stock. A Universal Cable Adaptor Recall Notification will be distributed by Certified Mail on or about November 30, 2004 providing instruction to remove the UCA and instruct consignees to follow November 1, 2004 recall interim instructions. Additionally, all consignees will receive instructions (on December 7, 2004) on how to return any device affected by the November 1, 2004 Recall for a power board replacement at a Pulmonetic systems, Inc. authorized service center.
Initiated
2004-11-19
Root cause
Other
Terminated
2005-12-13
Distribution
throughout the United States and to countries such as Germany, Canada, United Kingdom, Australia and Japan
Recall number
Z-0294-05
Recall status
Terminated
CDRH record ID
36028
FDA classified
2004-12-08
Recall event number
30514

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗