Reversed Instrument Position : a defect in the software version 3.0 causes an error in the position of surgical instruments and relative motion are reversed 180' on the StealthStation image. (If the StealthStation is not used, or if the patient orientation is prone or supine, the problem does not occur)
Source facts
Action
Medtronic Navigation notified customers by letter on October 17, 2007. Medtronic Field service will perform upgrade (software version 3.0.1) at user site.
Initiated
2007-10-17
Root cause
Software change control
Terminated
2007-11-27
Distribution
Worldwide: USA including states of CO, IN, NE, and MN and countries of Canada and Germany
Recall number
Z-0294-2008
Recall status
Terminated
CDRH record ID
65617
FDA classified
2007-11-27
Recall event number
45492