Skip to content

Public FDA record

Z-0294-2012: Health o Meter Professional Remote Display Digital Physician Scale, Model 498KL; digital patient weighing

Recall record Z-0294-2012 for Pelstar, Llc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0294-2012
Company in FDA record
Pelstar, Llc
FDA classification date
Nov 28, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

There is a potential for an electrical short when installing or replacing the batteries in the remote display head of the scales. This short may cause the batteries to smoke, burn or burst.

Source facts

Action
Pelstar LLC. sent a Urgent Device Recall letter dated October 7, 2011, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The letter informed them of the potential of an electrical short caused when installing or replacing the batteries which may result in minor injuries to the person installing the batteries. The customers were requested to immediately examine their inventory and quarantine all product subject to the recall. In addition, if they distributed this product to end users, they were requested to identify all customers and notify them of this product recall at once in writing, using the enclosed recall notification letter. The customers were also requested to complete and return the enclosed mandatory response form as soon as possible. Any questions were directed to Health O Meter Professional at 800-815-6615 or 708-377-0600.
Initiated
2011-10-07
Root cause
Device Design
Terminated
2017-11-30
Distribution
Worldwide Distribution -- USA (nationwide) including the countries of Canada, New Zealand, Israel and Lebanon.
Recall number
Z-0294-2012
Recall status
Terminated
CDRH record ID
104272
FDA classified
2011-11-28
Recall event number
60031

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗