Open original FDA record ↗Recall record
- FDA identifier
- Z-0294-2012
- Company in FDA record
- Pelstar, Llc
- FDA classification date
- Nov 28, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
There is a potential for an electrical short when installing or replacing the batteries in the remote display head of the scales. This short may cause the batteries to smoke, burn or burst.
Source facts
- Action
- Pelstar LLC. sent a Urgent Device Recall letter dated October 7, 2011, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The letter informed them of the potential of an electrical short caused when installing or replacing the batteries which may result in minor injuries to the person installing the batteries. The customers were requested to immediately examine their inventory and quarantine all product subject to the recall. In addition, if they distributed this product to end users, they were requested to identify all customers and notify them of this product recall at once in writing, using the enclosed recall notification letter. The customers were also requested to complete and return the enclosed mandatory response form as soon as possible. Any questions were directed to Health O Meter Professional at 800-815-6615 or 708-377-0600.
- Initiated
- 2011-10-07
- Root cause
- Device Design
- Terminated
- 2017-11-30
- Distribution
- Worldwide Distribution -- USA (nationwide) including the countries of Canada, New Zealand, Israel and Lebanon.
- Recall number
- Z-0294-2012
- Recall status
- Terminated
- CDRH record ID
- 104272
- FDA classified
- 2011-11-28
- Recall event number
- 60031