Open original FDA record ↗Recall record
- FDA identifier
- Z-0294-2013
- Company in FDA record
- MOOG Medical Devices Group
- FDA classification date
- Nov 15, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
MOOG Medical Devices Group is conducting a voluntary for a limited set of serial numbers of the EnteraLite Infinity and Infinity Orange Small Volume Enteral Feeding Pumps because certain diodes on the PCB assembly may result in insufficient voltage to a portion of the pump resulting in an error 13 message and shutting the pump down entirely.
Source facts
- Action
- MOOG Medical Devices Group sent a Urgent Medical Device Recall letter dated November 8, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to contact Moog Medical Devices Group to discuss the process for product replacement at 1-800-970-2337, prompt #7. See Attachment A to determine if you have a serial number that is related to a suspect pump. Determine the number of affected pumps. Notify all of your affected customers of the recall and the protocol for getting a replacement pump. Provide serial numbers, number of product(s) to be returned and a shipping address for replacement product. Execute the pump replacement process discussed in the conversation with Moog Customer Advocacy. Package and return product via shipping label.
- Initiated
- 2012-11-08
- Root cause
- Component design/selection
- Terminated
- 2014-01-30
- Distribution
- Worldwide Distribution - USA (nationwide) including the countries of Canada and Lithuania.
- Recall number
- Z-0294-2013
- Recall status
- Terminated
- CDRH record ID
- 114426
- FDA classified
- 2012-11-15
- Recall event number
- 63660