Z-0294-2013: EnteraLite Infinity Enteral Feeding Pump identified with the following list numbers: INFKIT2 and INFORNG.

Recall record Z-0294-2013 for MOOG Medical Devices Group. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0294-2013
Company in FDA record
MOOG Medical Devices Group
FDA classification date
Nov 15, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class II

Reason stated in the source

MOOG Medical Devices Group is conducting a voluntary for a limited set of serial numbers of the EnteraLite Infinity and Infinity Orange Small Volume Enteral Feeding Pumps because certain diodes on the PCB assembly may result in insufficient voltage to a portion of the pump resulting in an error 13 message and shutting the pump down entirely.

Source facts

Action
MOOG Medical Devices Group sent a Urgent Medical Device Recall letter dated November 8, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to contact Moog Medical Devices Group to discuss the process for product replacement at 1-800-970-2337, prompt #7. See Attachment A to determine if you have a serial number that is related to a suspect pump. Determine the number of affected pumps. Notify all of your affected customers of the recall and the protocol for getting a replacement pump. Provide serial numbers, number of product(s) to be returned and a shipping address for replacement product. Execute the pump replacement process discussed in the conversation with Moog Customer Advocacy. Package and return product via shipping label.
Initiated
2012-11-08
Root cause
Component design/selection
Terminated
2014-01-30
Distribution
Worldwide Distribution - USA (nationwide) including the countries of Canada and Lithuania.
Recall number
Z-0294-2013
Recall status
Terminated
CDRH record ID
114426
FDA classified
2012-11-15
Recall event number
63660

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗