Skip to content

Public FDA record

Z-0294-2017: Fusion Workstation.; Indicated for the transmission and review of radiological images.

Recall record Z-0294-2017 for Merge Healthcare, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0294-2017
Company in FDA record
Merge Healthcare, Inc.
FDA record creation date
Oct 20, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

After a period of time running Fusion Workstation, the Hounsfield measurement tool will report incorrect values.

Source facts

Action
The recalling firm initiated phone calls to affected sites beginning 1/30/2016.
Initiated
2011-10-28
Root cause
Software design
Terminated
2016-12-14
Distribution
Distribution was made to AZ, CA, ID, IL, IN, MI, MN, and NJ.
Recall number
Z-0294-2017
Recall status
Terminated
CDRH record ID
150241
FDA record created
2016-10-20
Recall event number
75425

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗