Z-0294-2018: LIFEPAK 20e DEFIBRILLATOR/MONITOR, 3201273 Product Usage: The LIFEPAK 20e defibrillator/monitor is inten

Recall record Z-0294-2018 for Physio-Control, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0294-2018
Company in FDA record
Physio-Control, Inc.
FDA classification date
Jan 11, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Physio-Control is aware that some devices have had power-related failures as customers prepared their device for initial deployment or during use within the first year of distribution. The symptoms of these failures may include unexpected power on and power off, device lock-up, or a failure to power on or off, any of which has the potential to result in a failure to deliver therapy to the patient and serious injury or death.

Source facts

Action
The recall was initiated by letter on 12/04/2017. The recall notice stated: "Physio-Control s Planned Actions Physio-Control will contact customers with LIFEPAK 20e devices who are potentially affected to arrange for a device correction. This correction will include the replacement of the Power PCBA. Required Customer Actions 1. Please forward this letter to all of your sites, trainers, and users that have an affected LIFEPAK 20e device(s) as identified on the attached Confirmation Sheet. 2. Promptly return the completed Confirmation Sheet to Physio-Control. 3. Follow the recommended daily Operator s Checklist steps in accordance with LIFEPAK 20e defibrillator/monitor Operating Instructions Section 7 Maintaining the Equipment. The checklist can be found in Appendix D of the Operating Instructions. 4. You may continue to use your defibrillator/monitor as long as it passes steps 1 and 8 of the Operator s Checklist. 5. If you experience any of the symptoms described above, or the defibrillator/monitor fails steps 1 and/or 8 of the Operator s Checklist, contact Physio-Control immediately to arrange servicing of your device. If your LIFEPAK 20e defibrillator/monitor exhibits any power issues that cannot be resolved, contact Physio-Control immediately. Contact Physio-Control if you have any questions about this matter at 1-800-442-1142, option 7, 6:00 A.M. to 4:00 P.M. (PST), Monday Friday."
Initiated
2017-12-04
Root cause
Other
Terminated
2024-04-01
Distribution
Worldwide Distribution
Recall number
Z-0294-2018
Recall status
Terminated
CDRH record ID
159792
FDA classified
2018-01-11
Recall event number
78574

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗