Reason stated in the source
Robust optimization for plans using beam set + background dose is not supported in RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6 and RayStation 7.
Source facts
- Action
- Ray Search Labs notified userson 9/21/18 by email advising users of the action to be taken: After creating an adaptive plan from a plan with robust objectives, remove all robust objectives that are automatically added to the adaptive plan and replace them with other appropriate objectives before optimizing the adapted plan. o Do not use an optimization function template with robust functions for a dependent beam set.o Do not use scripting to add robust optimization functions related to beam set + background dose. Reply form to be completed/signed and returned. Regulatory information, please contact quality@raysearchlabs.com
- Initiated
- 2018-09-21
- Root cause
- Software design
- Terminated
- 2020-06-16