Reason stated in the source

LIFEPAK 20 may lock-up when attempting to power-up on DC within 2 seconds after removing AC power.

Source facts

Action
Letter dated December 2006 alerts users to the lock-up problem, and suggesting a procedure to prevent lock-up. Firm promised another letter in January 2007 detailing the plan to provide software upgrade.
Initiated
2006-03-16
Root cause
Other
Terminated
2011-03-29
Distribution
Devices were distributed to hospitals and clinics worldwide.
Recall number
Z-0295-2007
Recall status
Terminated
CDRH record ID
48740
FDA classified
2006-12-19
Recall event number
36493

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗