Open original FDA record ↗Recall record
- FDA identifier
- Z-0295-2022
- Company in FDA record
- Medtronic Neuromodulation
- FDA record creation date
- Nov 24, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The wireless rechargers (WR; WR9200 and WR9220; included in the Recharger Kits) become unresponsive if the charging steps, indicated in the Recharger Patient User Guide, are not followed.
Source facts
- Action
- The recalling firm issued undated letters on 11/4/2021 via USPS Priority Shipping. The recalling firm has three different letters addressed to the Healthcare Professional depending upon the type of device the wireless recharger is used with (InterStim Micro, product number RS5200; Activa RC, product number RS6200; or RestoreUltra & RestoreSensor, Product number RS7200, which is currently not distributed in the U.S. at this time). The letter gives an issue overview and provides requested actions. Those actions include reinforcing the instructions in the enclosed Patient Quick User Guide with their patients and to attempt to reset the WR if the device appears unresponsive. If the problem is not resolved, then contact information is provided for Medtronic Technical Services. A customer confirmation form dated November 2021 was enclosed to acknowledge the consignee has received the letter.
- Initiated
- 2021-11-04
- Root cause
- Other
- Distribution
- US Nationwide, including PR. There was government distribution but no military distribution was identified.
- Recall number
- Z-0295-2022
- Recall status
- Open, Classified
- CDRH record ID
- 190097
- FDA record created
- 2021-11-24
- Recall event number
- 88953