Open original FDA record ↗Recall record
- FDA identifier
- Z-0297-2012
- Company in FDA record
- Microgenics Corp
- FDA classification date
- Nov 28, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Complaint from the sole distributor that the third party control material is recovering high, and outside the published range.
Source facts
- Action
- Microgenics sent a Product Removal notification letter dated June 15, 2011 to all affected customers. The letter identified the affected product, the problem, and the necessary actions to be taken. Customers were instructed to discontinue use and dispose of the affected product per local waste ordinances, review all patient results reported with the use of this lot, and retain a copy of this notification for their records. The letter states that if the affected product has been forwarded to another laboratory, a copy of the letter is to be provided to them. An alternative calibrator is available to customers impacted. It can be ordered by calling Thermo Fischer Scientific Customer Service at 1-800-232-3342. Questions regarding the information should be forwarded to Abbott Customer Service at 1-877-4ABBOTT.
- Initiated
- 2011-06-08
- Root cause
- Nonconforming Material/Component
- Terminated
- 2012-02-17
- Distribution
- Nationwide Distribution (USA) in the state of ILLINOIS
- Recall number
- Z-0297-2012
- Recall status
- Terminated
- CDRH record ID
- 104837
- FDA classified
- 2011-11-28
- Recall event number
- 60233