Skip to content

Public FDA record

Z-0297-2017: The PAXWIRE Occlusion Balloon System consists of an occlusion catheter, inflation handle and syringes. It

Recall record Z-0297-2017 for AccessClosure, Inc., A Cardinal Health Company. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0297-2017
Company in FDA record
AccessClosure, Inc., A Cardinal Health Company
FDA record creation date
Oct 21, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

AccessClosure, Inc. (ACI) is voluntarily recalling the PAXWIRE Occlusion Balloon System device because it was made aware from customers that the device may be removed in the absence of vascular sheath across the arteriotomy. This is a deviation from the firm's standard catheterization technique.

Source facts

Action
Cardinal Health sent an Urgent Medical Device Field Safety Notification letter to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed complete the attached Field Action Return Response Form as soon as possible indicating the number of devices that have been used as their location and the number that will be returned. Customers should discuss the reimbursement for their purchased and return product with their sales representative, who will coordinate the process. Customers with questions were instructed to call 408-610-6570. For questions regarding this recall call 408-610-6500.
Initiated
2016-10-06
Root cause
Device Design
Terminated
2016-11-14
Distribution
US distribution to FL and NJ
Recall number
Z-0297-2017
Recall status
Terminated
CDRH record ID
150167
FDA record created
2016-10-21
Recall event number
75376

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗