Skip to content

Public FDA record

Z-0297-2019: Multix Fusion system is a radiographic system: Models: Multix Fusion Analog (model number 10746665) with

Recall record Z-0297-2019 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0297-2019
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA record creation date
Oct 30, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Steel cables inside the lifting column of the overhead tube, may rapture without triggering the safety lock, leading to mechanical fatigue and cause the arm to drop down unexpectedly during movement in vertical direction and during patient positioning

Source facts

Action
Siemens Healthineers issued Urgent Medical Device Correction notice dated August 3, 2018, informing potentially affected users via Safety Advisory Notice XP029/18/S to advise of the problem, risk to health, action to be taken (it is strongly recommended not to perform up/down movements of the lifting column directly above the patient. Instead, the vertical movement should be completed beside the patient and then the system should be moved horizontally to the patient). The service organization will contact to arrange a date to perform this corrective action with Update Instruction XP031/18/S. Contact service organization for an earlier appointment at 1-800-888-7436. Siemens Healthineers is preparing a corrective action that will resolve this potential issue. The corrective action will be distributed via an Update Instruction XP031/18/S.
Initiated
2018-08-03
Root cause
Device Design
Terminated
2019-11-26
Distribution
US nationwide distribution
Recall number
Z-0297-2019
Recall status
Terminated
CDRH record ID
167219
FDA record created
2018-10-30
Recall event number
81007

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗