Open original FDA record ↗Recall record
- FDA identifier
- Z-0297-2019
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA record creation date
- Oct 30, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Steel cables inside the lifting column of the overhead tube, may rapture without triggering the safety lock, leading to mechanical fatigue and cause the arm to drop down unexpectedly during movement in vertical direction and during patient positioning
Source facts
- Action
- Siemens Healthineers issued Urgent Medical Device Correction notice dated August 3, 2018, informing potentially affected users via Safety Advisory Notice XP029/18/S to advise of the problem, risk to health, action to be taken (it is strongly recommended not to perform up/down movements of the lifting column directly above the patient. Instead, the vertical movement should be completed beside the patient and then the system should be moved horizontally to the patient). The service organization will contact to arrange a date to perform this corrective action with Update Instruction XP031/18/S. Contact service organization for an earlier appointment at 1-800-888-7436. Siemens Healthineers is preparing a corrective action that will resolve this potential issue. The corrective action will be distributed via an Update Instruction XP031/18/S.
- Initiated
- 2018-08-03
- Root cause
- Device Design
- Terminated
- 2019-11-26
- Distribution
- US nationwide distribution
- Recall number
- Z-0297-2019
- Recall status
- Terminated
- CDRH record ID
- 167219
- FDA record created
- 2018-10-30
- Recall event number
- 81007