Skip to content

Public FDA record

Z-0298-2024: LIFEPAK 15, REF: 99577-000061, Part Number V15-2-000067

Recall record Z-0298-2024 for Physio-Control, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0298-2024
Company in FDA record
Physio-Control, Inc.
FDA classification date
Nov 14, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Their is a potential for battery pins to become loose within the defibrillator resulting in power issues.

Source facts

Action
On October 6, 2023, Stryker issued a "Urgent Medical Device Correction" Notification to affected consignees. Stryker ask consignees to take the following actions: 1. Please check your internal inventory to locate the affected product listed on the attached business reply form. 2. To confirm receipt of this Medical Device Notice and understanding of the provided information, please email the enclosed Business Reply Form (BRF) by October 31, 2023, to RSRecall@Stryker.com. 3. Once the BRF is received, a Stryker Representative will contact you to arrange for corrective action on the device. There are no additional actions that users can and should take. 5. In the interim, Stryker recommends that you continue to use your LP15 v4 device. 6. Maintain awareness of this communication internally until all required actions have been completed within your facility. 7. Inform Stryker if any of the subject devices have been distributed to other organizations by contacting Stryker at: a. Device Tracking: rsdevicetracking@stryker.com b. Website: https://www.strykeremergencycare.com/service-support-overview/device-registration/
Initiated
2023-09-29
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution in the states of AZ, CA, NC, NJ, NY.
Recall number
Z-0298-2024
Recall status
Open, Classified
CDRH record ID
203905
FDA classified
2023-11-14
Recall event number
93233

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗