Skip to content

Public FDA record

Z-0299-2009: Zimmer Alumina Ceramic Femoral Head, 12/14 taper, 28 mm diameter, neck length -3.5 mm, sterile, Zimmer, W

Recall record Z-0299-2009 for Zimmer Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0299-2009
Company in FDA record
Zimmer Inc.
FDA classification date
Nov 21, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

When used with cobalt/chromium hip stems, the average test values for ceramic head bursting falls below the guidance document limits if the taper is made by casting instead of by forging.

Source facts

Action
The firm issued Product Compatibility Update # 13, dated 6/11/08, to consignees informing them which products are not compatible, and thus should not be used in combination.
Initiated
2008-06-11
Root cause
Device Design
Terminated
2009-11-10
Distribution
Worldwide Distribution --- including USA and countries of Austria, Australia, Belgium, Brazil, Canada, Chile, China, Czech Republic, Denmark, Dubai, England, Finland, France, Germany, Greece, Hungary, India, Israel, Italy, Korea, Lebanon, Mexico, Netherlands, New Zealand, Poland, Romania, Singapore, Slovakia, Spain, Switzerland and Ukraine.
Recall number
Z-0299-2009
Recall status
Terminated
CDRH record ID
71981
FDA classified
2008-11-21
Recall event number
48712

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗