Z-0300-06: Titanium Greenfield Vena Cava Filter with 12 Fr./4,0 mm Introducer System. Catalog no. M001503010. Ster

Recall record Z-0300-06 for Boston Scientific Scimed. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0300-06
Company in FDA record
Boston Scientific Scimed
FDA classification date
Dec 23, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Certain units within eleven lots of the 12 Fr Greenfield Vena Cava Filter may lack a taper on the braided sheath of the delivery system. The lack of a taper could lead to vessel damage as a result of the sheath catching on the edge of the vein and causing a tear.

Source facts

Action
Urgent Recall Notice dated 08/04/05 was Federal Express''d to customers for delivery on 08/05/05. This notice informs customers of the issue and identifies affected product. Customers are asked to immediately discontinue use of and segregate these recalled lots. Steps for Voluntary Recall and Tracking/Verification Form were sent to the Risk Manager only. The Tracking/Verification Form is to be completed by the Risk Manager even if there are no affected lots to return. To return affected product, the customer must call Boston Scientific Corporation (BSC) to obtain a Returned Goods Authorization (RGA) number. Returned Goods are to be packaged and sent to BSC with the Tracking/Verification Form and using the enclosed red/white field action return shipping label to: Boston Scientific Marina Bay, 500 Commander Shea Blvd., Quincy, MA 02171
Initiated
2005-08-05
Root cause
Other
Terminated
2006-11-09
Distribution
OUS includes countries of Chile, Japan, Netherlands and Taiwan. Within the US to: AK, AL, AR, AZ, CA, CO, CT, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, NC, NE, NH, NJ, NY, OH, OR, PA, SC, TN, TX, VA and WI
Recall number
Z-0300-06
Recall status
Terminated
CDRH record ID
41033
FDA classified
2005-12-23
Recall event number
32889

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗