Reason stated in the source
An error occurred at the board manufacturing site which may affect the following parameters: - An electrical noise observed on the SpO2 pleth waveform generating and displaying erratic heart rates. % SpO2 reading displays dashes. - The ECG waveform may appear noisy, ECG readings appear accurate. -Respiration waveform may appear noisy and the respiration readings will be displayed as dashes.
Source facts
- Action
- An Safety Action Bulletin, dated 11/03/2006 was sent to affected customers describing the issue, what parameters are affected on the monitor and provides informatin on how to have the monitor repaired. Smiths Medical MDPM requests that an acknowledgement form be returned to them.