Z-0300-2008: Ventana Medical System's Symphony Staining System, KPA, Model #: 2200000, Catalog #: 900-SYM1 & 900-SYM3,

Recall record Z-0300-2008 for Ventana Medical Systems Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0300-2008
Company in FDA record
Ventana Medical Systems Inc
FDA classification date
Jan 9, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Shock hazard: Devices released prior to full testing; the configuration of the Symphony instrument during Hi pot testing procedures did not completely evaluate all components in the test loop. In specific circumstances, a shock hazard exists.

Source facts

Action
On 9/6/07, the firm began visiting customers to perform the HiPot test. Prior to performing the High Potential Test of Symphony Instruments already in the Field, the Field Service Engineer or FSE ensures that the customer fully understands the purpose of the test and why it is necessary to perform the test at their site. The FSE conveys the following points regarding the HiPot test: - The HiPot test is a safety verification for the electrical path within the instrument. - The HiPot test was performed incorrectly during the manufacturing process. - The HiPot test is used to verify that if a high voltage overload condition were to occur it would not lead to a short/arc within the instrument, which could pose a risk to anyone that touches the instrument - The HiPot test places a voltage of 2,000VDC onto the high voltage path of the instrument and if any of the insulating components are compromised, the tester will notify the FSE of the condition. - If the instrument fails the HiPot test, the faulty component will be replaced and the instrument will be retested until it passes. - Due to the high voltage used the during the test, no one will be allowed to touch the instrument until the test is complete. The FSE performing the test documents that the customer was informed of the reason for the High Potential Test on their Field Service Report.
Initiated
2007-09-06
Root cause
Process control
Terminated
2008-07-08
Distribution
Worldwide: USA, Japan and France
Recall number
Z-0300-2008
Recall status
Terminated
CDRH record ID
65469
FDA classified
2008-01-09
Recall event number
45437

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗