Z-0301-2019: Roche Diagnostics URISYS 1100 Urine Analyzer Part Number: 03617556001 Product Usage: The URISYS 1100 Uri

Recall record Z-0301-2019 for Roche Diagnostics Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0301-2019
Company in FDA record
Roche Diagnostics Corporation
FDA record creation date
Oct 31, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Limits of Detection (LoD) for protein, nitrite, leukocytes, and erythrocytes on the Urisys 1100 urine analyzer with Chemstrip 5 OB, Chemstrip 7, Chemstrip 10 MD, and Chemstrip 10 UA test strips were higher than what is listed in the test strip method sheets may lead to false negative results for the affected parameters

Source facts

Action
Roche Diagnostics notified Consignees by the Urgent Medical Device Correction (UMDC) sent UPS Ground (receipt required) on 9/24/18. For consignees associated only with a PO number address, the UMDC will be sent by USPS mail. Consignee Instructions: 1. When using Chemstrip 5 Ob and/or Chemstrip 7 test strips, results should only be read visually. These test strips can no longer be read on the Urisys 1100 urine analyzer. Follow the directions in the Instructions for Manually Reading Chemstrip test Strip results section of the UMDC. 2. When using Chemistrip 10 MD or Chemstrip 10 UA test strips: You can choose to read the test strips visually or If you continue to use the strips with the Urisys 1100 urine analyzer, a negative result for any of the four affected parameters (protein, nitrite, leukocytes, and erythrocytes) must be repeated with a new test strip that is read visually. Follow the directions in the instructions for Manually Reading Chemstrip Test Strip Results section of the UMDC. 3. If you are unable to implement the visual read-only method for Chemstrip 5 OB, Chemstrip 7, Chemstrip 10 MD, or Chemstrip 10 UA test strips: Dispose of the affected test strip vials per local guidelines. Contact your distributor partner for credit. 4. Complete all sections of the enclosed fax back forms Contact Roche Diagnostics Point-of-Care Technical Service, 24 hours a day, seven days a week at 1-800-428-4674 if you have questions
Initiated
2018-09-24
Root cause
Under Investigation by firm
Terminated
2022-02-09
Distribution
US Nationwide Distribution
Recall number
Z-0301-2019
Recall status
Terminated
CDRH record ID
167677
FDA record created
2018-10-31
Recall event number
81177

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗