Z-0301-2021: 5.5MM Arthrogarde Hip Access Cannula

Recall record Z-0301-2021 for Smith & Nephew, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0301-2021
Company in FDA record
Smith & Nephew, Inc.
FDA record creation date
Oct 27, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The J Lock feature on the cannula body was manufactured incorrectly, preventing the cannula from locking into an irrigation extender or arthroscope in the correct orientation.

Source facts

Action
On 9/21/2020 the firm send a notification to its consignees with the following instructions: Required Actions: 1. Please inspect your inventory and locate any devices from the listed product and batch numbers enclosed with this Urgent Medical Device Recall Notice, and quarantine them immediately. a. If you are a sales rep or distributor, you must notify your customers of the field action and ensure that the required actions are complete. b. Include the customers account number and name in the table below. 2. If you have no product to return, please put an X in the appropriate location below. 3. If you have product to return, please list the item, batches and quantities that you are returning in the appropriate boxes below. 4. Complete the remainder of the form, sign and send to FieldActions@smith-nephew.com. Please Note even if you have no product to return, this form must be completed, signed and returned. 5. Once the form is received by Smith & Nephew, you will be sent a Return Authorization (RA) number.
Initiated
2020-09-21
Root cause
Process control
Terminated
2024-03-22
Distribution
Domestic Distribution: CA, CO, CT, GA, IA, IL, IN, MD, MO, NC, NY, OH, PA, SC, TN, TX, WA, WI, WV. International Distribution: AU, BR, CA, CL, CN, CO, DE, ES, GB, HK, KR, PA, PT, TW.
Recall number
Z-0301-2021
Recall status
Terminated
CDRH record ID
183839
FDA record created
2020-10-27
Recall event number
86502

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗