Skip to content

Public FDA record

Z-0301-2027: EMAG system Model/Catalog Number: 418591/418591U Product Description: Nucleic acid sample preparation ins

Recall record Z-0301-2027 for BioMerieux SA. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0301-2027
Company in FDA record
BioMerieux SA
FDA classification date
Oct 27, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Delayed or no result when analyzing clinical samples with EMAG system (ref: 418591/418591U) due to a randomly occurring software error, Error 10003 Heater timeout after updating to recent firmware SP4 Patch 2

Source facts

Action
The firm began notifying consignees on September 11, 2026. Customers were informed they may continue use of EMAG systems and there is no risk of false results. In case of 10003 error heater time-out during an extraction protocol, customers are instructed to reset the system to run the protocol again to prevent delay/loss of sample. The company will contact customers to schedule a bioMerieux Field Service Engineer (FSE) visit to update affected EMAG systems to the previous firmware patch SP4 Patch 1 to fix the issue.
Initiated
2026-09-11
Root cause
Software change control
Distribution
Worldwide - US Nationwide distribution in the states of CA, IN, KS, LA, MN, MO, NY, OH, and TX. The countries of Canada, United Kingdom, Sweden, France, Norway, Netherlands, Belgium, Denmark, Germany, Austria.
Recall number
Z-0301-2027
Recall status
Open, Classified
CDRH record ID
222883
FDA classified
2026-10-27
Recall event number
99931

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗