Software Correction: Software errors that may result in incorrect 4D CT images, PET images with SUV quantification error, or disruption of interventional CT procedures; and Software errors that may result in partial set of images, inability to generate CT or PET images, incorrect scan parameters, or scanning the incorrect portion of the body.
Source facts
Action
Philips notified customers of these issues through Customer Information Letter (CIL) 88200518 to U.S. customers on 9/04/18 receive via USPS Certified mail.
notifying them about the issues and temporary solutions available.
Non-U.S. customers Non-U.S. customers will receive the letter through Philips Global Markets following local regulations.
Field correction will consist of proactive field change order (FCO) release to be deployed to the impacted systems to correct all these issues.customers.
if you need any further information or support concerning this issue, contact your local Philips representative.
For North America and Canada, contact the Customer Care Solutions Center (1-
800-722-9377, follow the prompts).
Initiated
2018-09-04
Root cause
Software design
Terminated
2020-05-05
Distribution
US Nationwide Distribution
Recall number
Z-0302-2019
Recall status
Terminated
CDRH record ID
167661
FDA record created
2018-10-31
Recall event number
81173