Open original FDA record ↗Recall record
- FDA identifier
- Z-0302-2022
- Company in FDA record
- Siemens Medical Solutions USA, Inc
- FDA record creation date
- Nov 24, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
If SID (source-to-image distance) lift movement is activated and x-ray shall be released simultaneously; x-ray is not possible and the message No x-ray, try again is displayed, may result in a short delay in procedure
Source facts
- Action
- Siemens Medical Solutions USA, Inc. initiated a Customer Safety Advisory Notice dated 10/202/1 to customers via AX070/21/S, Siemens will update the system software via Update Instruction AX069/21/S. Letter states reason for recall, health risk and action to take: To avoid the problem, it is recommended to position the SID first and then release x-ray. Siemens will correct the software error (software VD12) via Update Instruction AX069/21/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. We do not consider it necessary to re-examine any patients in relation to the issues described above. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content therein
- Initiated
- 2021-10-20
- Root cause
- Software design
- Terminated
- 2024-06-18
- Distribution
- US Nationwide Distribution
- Recall number
- Z-0302-2022
- Recall status
- Terminated
- CDRH record ID
- 190135
- FDA record created
- 2021-11-24
- Recall event number
- 88962