Skip to content

Public FDA record

Z-0303-2017: ReTurn 7500/7500 Aid, transfer

Recall record Z-0303-2017 for Handicare USA, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0303-2017
Company in FDA record
Handicare USA, Inc.
FDA record creation date
Oct 24, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During the period of April 2009 May 2011, Handicare AB used a wing handle in which the screw and handle, after repeated tightening, may slide apart. This causes a small amount of play in the attachment of the ladder, which in turn results in high load at the attachment holes of the ladder. Eventually, this may cause mechanical fatigue in the material of the ladder.

Source facts

Action
Handicare sent a Field Safety notice dated April 21, 2014, to all affected customers. The notification instructed customers to examine the affected ReTurn units for signs of wear and to complete the response form that was mailed with the customer notification letter indicating whether additional parts were required to complete the field correction. For further questions, please call (717) 733-1444.
Initiated
2014-04-15
Root cause
Nonconforming Material/Component
Terminated
2016-12-13
Distribution
US Distribution to the states of : NY, NJ, NV, NE, IL, CT, PA and ME
Recall number
Z-0303-2017
Recall status
Terminated
CDRH record ID
150211
FDA record created
2016-10-24
Recall event number
75404

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗