Open original FDA record ↗Recall record
- FDA identifier
- Z-0303-2017
- Company in FDA record
- Handicare USA, Inc.
- FDA record creation date
- Oct 24, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
During the period of April 2009 May 2011, Handicare AB used a wing handle in which the screw and handle, after repeated tightening, may slide apart. This causes a small amount of play in the attachment of the ladder, which in turn results in high load at the attachment holes of the ladder. Eventually, this may cause mechanical fatigue in the material of the ladder.
Source facts
- Action
- Handicare sent a Field Safety notice dated April 21, 2014, to all affected customers. The notification instructed customers to examine the affected ReTurn units for signs of wear and to complete the response form that was mailed with the customer notification letter indicating whether additional parts were required to complete the field correction. For further questions, please call (717) 733-1444.
- Initiated
- 2014-04-15
- Root cause
- Nonconforming Material/Component
- Terminated
- 2016-12-13
- Distribution
- US Distribution to the states of : NY, NJ, NV, NE, IL, CT, PA and ME
- Recall number
- Z-0303-2017
- Recall status
- Terminated
- CDRH record ID
- 150211
- FDA record created
- 2016-10-24
- Recall event number
- 75404