Open original FDA record ↗Recall record
- FDA identifier
- Z-0304-2008
- Company in FDA record
- Siemens Medical Solutions USA, Inc.
- FDA classification date
- Dec 19, 2007
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Unintended Motion: A hardware fault in the c.cam system can cause the system to experience unintended motion, thereby causing a collision hazard.
Source facts
- Action
- Urgent Field Correction Recall letters dated 9/11/07, on Siemens letter head were sent to the end users via U.S. certified mail, informing them of the potential risk of unintended motion of the detector head resulting in a collision hazard. The end users were warned not to press any of the detection motion buttons while simultaneously holding the collision over ride button on the c.cam hand controller, and were provided steps to take if the system stops during an acquisition and displays the Collision Error message. They were told that a Siemens service representative will contact them within 90 days to schedule the repair of the c.cam at no cost to the end user. Any questions regarding the Important Safety Notice were directed to their service representative or Siemens office at: 1-800-888-7436 (USA).
- Initiated
- 2007-09-17
- Root cause
- Software design
- Terminated
- 2009-01-09
- Distribution
- Worldwide: USA including Puerto Rico and Brazil, Canada, France, Germany, Great Britain, Israel, The Netherlands, and Portugal.
- Recall number
- Z-0304-2008
- Recall status
- Terminated
- CDRH record ID
- 64526
- FDA classified
- 2007-12-19
- Recall event number
- 44913