Open original FDA record ↗Recall record
- FDA identifier
- Z-0304-2012
- Company in FDA record
- Del Mar Reynolds Medical, Ltd.
- FDA classification date
- Nov 29, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
If external pressure is applied to the cuff of the Ambulatory Blood Pressure, Model 90217A, during deflation, the monitor will stop working and display an EC75 error code.
Source facts
- Action
- On October 21, 2011, a Medial Device Correction customer letter was mailed to all US customers. A return receipt was requested. The letter identified the affected devices and described the problem. If the device displays an EC75 error code, customers can cycle the power off and then on to restore normal functionality. Please advise patients of this mitigation to avoid a delay in obtaining results. The firm is in the process of upgrading their software to avoid the issue and will contact customers to arrange for an update to their monitors. Customers can call Spacelabs Healthcare at 1-800-522-7025, select 2 for Technical support for any question about this recall.
- Initiated
- 2011-10-12
- Root cause
- Device Design
- Terminated
- 2012-05-30
- Distribution
- Worldwide Distribution
- Recall number
- Z-0304-2012
- Recall status
- Terminated
- CDRH record ID
- 104737
- FDA classified
- 2011-11-29
- Recall event number
- 60204