Open original FDA record ↗Recall record
- FDA identifier
- Z-0304-2017
- Company in FDA record
- Nova Ortho-Med Inc
- FDA classification date
- Oct 24, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Nova Medical is recalling the Journey Rolling Walker because the screws used to hold the black side support pieces to the frame can become loose over time. This can result in the locking nut coming off and the screw falling out.
Source facts
- Action
- A customer notification letter dated 10/1/16 was sent to customers to inform them that the screws used to hold the black side support pieces to the frame can become loose over time. This can result in the locking nut coming off and the screw falling out. This allows the frame support piece to detach from the walker and causes instability resulting in possible injury. The letter informs the customers of the root cause and the solution. Customers are instructed to record the serial #s of the product that have been corrected and email them to NOVA Customer Service (patricias@novajoy.com). Customers can also 1-800-557-6682 x131.
- Initiated
- 2016-10-01
- Root cause
- Component design/selection
- Terminated
- 2017-02-06
- Distribution
- US Nationwide Distribution in the states of: CA, IN, HI, CT, UT, WA, FL, VA, WI, OH, MN, KS, MD, IL, MO, WY, MT, ME, PA, TN, MI, OR, NY, NJ, ID, NC, IA, OK, AR, VT, SC, TX
- Recall number
- Z-0304-2017
- Recall status
- Terminated
- CDRH record ID
- 150220
- FDA classified
- 2016-10-24
- Recall event number
- 75408