Open original FDA record ↗Recall record
- FDA identifier
- Z-0305-2008
- Company in FDA record
- Toshiba American Med Sys Inc
- FDA classification date
- Jan 9, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Panel Lock-up: When using dynamic image acquisition modes the panel may lock up, if the operator attempts to perform a still image capture. If the panel locks up the system will have to be rebooted to continue the examination which may cause a delay in completion of the patient examination.
Source facts
- Action
- Notification letters, Urgent Medical Device Correction, with an attached Customer Response form were sent to customers beginning on October 2, 2007. Customers were notified this action was being taken as a result of an investigation that resulted from a software issue that was found by the manufacturer. Customers were informed that a Toshiba Customer Service Engineer would contact them to schedule a software upgrade when the software becomes available. If customers have any questions they were instructed to contact their or call (800) 421-1968 and ask for the Regulatory Affairs Department.--A short term instruction included "Avoid using the Freeze Key when performing dynamic image acquisitions"
- Initiated
- 2007-10-02
- Root cause
- Software design
- Terminated
- 2012-04-20
- Distribution
- Nationwide: USA including the states of AK, AZ, CA, FL, IA, IL, IN, KS, MI, MN, MO, MT, NE, NV, NY, OH, OR, PR, VA, WI & WV
- Recall number
- Z-0305-2008
- Recall status
- Terminated
- CDRH record ID
- 66198
- FDA classified
- 2008-01-09
- Recall event number
- 45811