Skip to content

Public FDA record

Z-0305-2017: Symbiq One Channel Infuser; an Rx medical device infusion pump used to administer I.V. fluids Device is

Recall record Z-0305-2017 for Hospira Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0305-2017
Company in FDA record
Hospira Inc.
FDA record creation date
Oct 24, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Symbiq Infusers have the potential to experience a white screen during titration of a Keep Vein Open (KVO) delivery. This can allow the clinician to select the "NEXT" button prior to completion of the programming. If this error occurs, a delay in therapy could occur and potentially result in significant injury.

Source facts

Action
Hospira sent an Urgent Device Field Correction notice dated February 4, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to follow the steps provided in the letter while Hospira is developing a software change. Customers were asked to complete the attached reply form and return it via fax to the number on the form. Customers with questions were instructed to contact Stericycle at 1-888-201-9332. For questions regarding this recall call 224-212-4892.
Initiated
2013-02-04
Root cause
Software design
Terminated
2016-12-15
Distribution
Worldwide Distribution - USA (nationwide) and Canada
Recall number
Z-0305-2017
Recall status
Terminated
CDRH record ID
115185
FDA record created
2016-10-24
Recall event number
63994

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗