Skip to content

Public FDA record

Z-0305-2021: CMEAmerica BodyGuard Wall Charger: Lot number 2019-0364, Mode/Catalogue number 151-143XL, *+B 01151143X L

Recall record Z-0305-2021 for CME America LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0305-2021
Company in FDA record
CME America LLC
FDA record creation date
Oct 29, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The wall charger may not properly charge the infusion pump battery.

Source facts

Action
On February 27, 2020, the firm sent an "URGENT MEDICAL DEVICE RECALL" of wall charger, not charging the infusion pump, which could result in an unexpected low battery power. This could result in delay of medications and infusions. The firm instructed their customers to review their inventory.The firm will initiate contact with the customers via email and written letter by FedEx. A response form will be included for customer acknowledgement. The firm will request that distributors provide their customer list and the firm will notify those customers via FedEx mail. A customer response form is included with the letter to assist consignees in responding to the field action. The response forms will be reconciled with the customer list. The firm will make at least three attempts to follow up with all non-respondents via telephone, email or mail. The firm instructed their customers of the following actions: 1.Immediately review their inventory for the recalled devices, check the functionality of wall chargers, which displayed red light if functioning. Discard all non-functional wall chargers 2. Share this Urgent Medical Device Recall notice with all users of the product within your facility to ensure awareness of this action. 4.Complete the attached Customer Response Form and return it to the contact noted on the form regardless of whether the customers have any affected wall chargers or not so that the firm can acknowledge the receipt of this notification. 5. The firm asked customers to indicate on the response form the quantity of affected product identified at their facility and confirm that this inventory has been discarded. Credit or replacement product will be issued to customers for all discarded product. The firm will provide credit or replacements for all discarded inventory and provided the contact information as: Customer Support 303-936-4545 or 877-263-0111 toll free
Initiated
2020-02-27
Root cause
Employee error
Terminated
2022-01-26
Distribution
U.S. Consignees: 3, IL, MO, KS OUS : 0 "No foreign consignees".
Recall number
Z-0305-2021
Recall status
Terminated
CDRH record ID
181046
FDA record created
2020-10-29
Recall event number
85479

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗