- Action
- On September 30, 2021, the firm sent out a customer notification letter "Urgent Medical Device Recall" to their customers via Fed-Ex with the instruction due to the packaging breach, identified during the in house internal leak testing. This breach in the product packaging rendered the enclosed IV catheters non-sterile. Holes have been detected on the bottom tray of device unit packages. The notification stated to destroy the devices which are present in customers/distributors inventory. The recalling firm notified in a customer notification letter to fill and send the customer response form with customer's details and acknowledgement of destruction of the device list , lot number and quantities for the replacement. The contact information will be provided for the customers to call : 1-844-8BD-LIFE (1-844-823-5433)Say Recall when prompted, Mon Fri 8:00am and 5:00pm CT
or Email: productcomplaints@bd.com
The customer response form to be returned via: E-Mail BDRC17@bd.com, Fax No. is 312-949-0375.
- Initiated
- 2021-09-08
- Root cause
- Packaging process control
- Distribution
- U.S. Nationwide Distribution: AR, AZ, MA, GA, NC, OH, WA, NY, NJ, HI, CA, MD, RI, MT, IN, PA, IL OK, OR, AK, CT, TX, ME, MO, FL, VA, LA, CO, UT
O.U.S.: None
- Recall number
- Z-0306-2022
- Recall status
- Open, Classified
- CDRH record ID
- 189467
- FDA record created
- 2021-11-24
- Recall event number
- 88762