Skip to content

Public FDA record

Z-0306-2022: BD Nexiva, 20 GA 1.00 IN (1.1 X 25 mm), 3660 ml/hr. 61ml/min; Closed IV Catheter System- Dual port -61 mL

Recall record Z-0306-2022 for Becton Dickinson Infusion Therapy Systems Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0306-2022
Company in FDA record
Becton Dickinson Infusion Therapy Systems Inc.
FDA record creation date
Nov 24, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a breach in the product packaging that renders the product non-sterile.

Source facts

Action
On September 30, 2021, the firm sent out a customer notification letter "Urgent Medical Device Recall" to their customers via Fed-Ex with the instruction due to the packaging breach, identified during the in house internal leak testing. This breach in the product packaging rendered the enclosed IV catheters non-sterile. Holes have been detected on the bottom tray of device unit packages. The notification stated to destroy the devices which are present in customers/distributors inventory. The recalling firm notified in a customer notification letter to fill and send the customer response form with customer's details and acknowledgement of destruction of the device list , lot number and quantities for the replacement. The contact information will be provided for the customers to call : 1-844-8BD-LIFE (1-844-823-5433)Say Recall when prompted, Mon Fri 8:00am and 5:00pm CT or Email: productcomplaints@bd.com The customer response form to be returned via: E-Mail BDRC17@bd.com, Fax No. is 312-949-0375.
Initiated
2021-09-08
Root cause
Packaging process control
Distribution
U.S. Nationwide Distribution: AR, AZ, MA, GA, NC, OH, WA, NY, NJ, HI, CA, MD, RI, MT, IN, PA, IL OK, OR, AK, CT, TX, ME, MO, FL, VA, LA, CO, UT O.U.S.: None
Recall number
Z-0306-2022
Recall status
Open, Classified
CDRH record ID
189467
FDA record created
2021-11-24
Recall event number
88762

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.