Z-0307-2011: WalkMed Triton Pole Mount Infusion Pump, Catalog 300000. Indicated for delivering infusions of medicatio

Recall record Z-0307-2011 for Walkmed Infusion LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0307-2011
Company in FDA record
Walkmed Infusion LLC
FDA classification date
Nov 5, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Pump door may be in a near shut position, but unlatched, and the "Door Open" alarm may not sound.

Source facts

Action
All customers were notified by letter on 06/17/2010, describing the affected product and advising them of the possible malfunctioning alarm in the case a user does not follow the instructions as specified in the Triton Operation Manual. The letter also reminded all users to follow the procedures in the IFUs. The firm stated that it was best if customers re-educated all nurses on the proper door closure procedure to help prevent the potential bypass issue from occurring. Attached to the letter were two pages explaining the proper process for closing the pump door. Follow-up phone calls were made to coordinate an upgrade to all affected pumps. Questions or concerns should be directed to Ross Kurz at 720-531-4933.
Initiated
2010-06-17
Root cause
Device Design
Terminated
2012-03-14
Distribution
Nationwide Distribution -- AL, CA, FL, and PA.
Recall number
Z-0307-2011
Recall status
Terminated
CDRH record ID
94445
FDA classified
2010-11-05
Recall event number
56750

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗