Open original FDA record ↗Recall record
- FDA identifier
- Z-0307-2012
- Company in FDA record
- Elekta, Inc.
- FDA classification date
- Nov 30, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
There is a risk that the user may interpret the instructions in the manual too literally, and adjust the TMR 10 dose description to obtain an ETCPD that matches previous clinical prescriptions, which in turn could lead to a prescription dose that deviates a few percent from the intended dose.
Source facts
- Action
- Elekta sent an IMPORTANT NOTICE letter dated June 23, 2011 to affected customers. The notice identified the product, problem, and corrective actions to be taken by the user. The letter advise all customers to follow any instructions or recommendations covered in the Notice. The letter states that Elekta is working on a new software version and customers will receive the future version when released. For questions contact your local Elekta representative.
- Initiated
- 2011-06-23
- Root cause
- Labeling False and Misleading
- Terminated
- 2012-01-31
- Distribution
- Nationwide Distribution (USA) - including the states of: NM and WA
- Recall number
- Z-0307-2012
- Recall status
- Terminated
- CDRH record ID
- 104476
- FDA classified
- 2011-11-30
- Recall event number
- 60125