Skip to content

Public FDA record

Z-0307-2027: PENTAX Medical Ultrasound Endoscope. Model Numbers. EG38-J10UT, EG34-J10U.

Recall record Z-0307-2027 for Pentax of America Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0307-2027
Company in FDA record
Pentax of America Inc
FDA classification date
Oct 28, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Warnings added to the IFU for ultrasound endoscopes and water bottle assembly to ensure there is deaerated sterile water in the assembly prior to and during the procedure.

Source facts

Action
An Urgent Field Safety Notice dated 09/02/2026 was mailed to consignees. The notice informs consignees of two updates to the IFU of PENTAX Medical products, specifically the ultrasound endoscope devices and water bottle assembly. Consignees are to inform their staff of this notice and include the warning and updates to their IFUs. Additionally, consignees are instructed to change water for each procedure and fill the bottle approximately 2/3 full, or to the 200 marking on the bottle scale, with sterile water. If the water/balloon feeding function does not work consignees are to fill the water bottle with deaerated sterile water. PENTAX Medical can be contacted by phone at 1-800-431-5880 Monday through Friday from 8:30 AM EST to 5:00 PM EST, or by email at customeradvisories@pentaxmedical.com.
Initiated
2026-09-09
Root cause
Unknown/Undetermined by firm
Distribution
Worldwide - US Nationwide and International distribution - Pending.
Recall number
Z-0307-2027
Recall status
Open, Classified
CDRH record ID
223088
FDA classified
2026-10-28
Recall event number
99782

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗