Potential for shock treatment to be delayed or prevented due to board contamination that causes intermittent switch operation
Source facts
Action
On 12/19/06 the firm sent letters via return receipt to all domestic customers. The letter advised customers of the issue and if the user encounters the problem, to press the button again with additional force. The users are to contact Philips for replacement. The devices can be used until the user receives a replacement.
Initiated
2006-12-19
Root cause
Process design
Terminated
2011-06-22
Distribution
Distribution is worldwide.
Recall number
Z-0308-2007
Recall status
Terminated
CDRH record ID
49112
FDA classified
2006-12-27
Recall event number
36662