Z-0308-2012: ViroMax viral/bacterial filter. The filter reduces the risk of infection due to cross-contamination betwe

Recall record Z-0308-2012 for A M Systems Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0308-2012
Company in FDA record
A M Systems Inc
FDA classification date
Jan 11, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Some ViroMax viral/bacterial filters are leaking at the point where the two halves of the filter are ultrasonically bonded together.

Source facts

Action
A - M Systems sent a Urgent Notification of Product Recall/Correction dated March 31, 2010, to all affected customers. The letter identificed the product the problem and the action needed to be taken by the customer. Customers were instructed to cease use and/or distribution of the affected product and to immediately notify any of their sub-accounts with the same information. Customers are asked to provide A - M Systems with the amount distributed and amount remaining in inventory and to then destroy or discard any remaining product in their possession. Please call 360.683.8300, if you have any questions or concerns.
Initiated
2010-03-31
Root cause
Nonconforming Material/Component
Terminated
2012-01-05
Distribution
Worldwide Distribution -- USA ( nationwide) including the states of AZ, CA, FL, KY, MA, MI, NC, OH, OR, and WI., and the countries Canada, Mexico, and Taiwan.
Recall number
Z-0308-2012
Recall status
Terminated
CDRH record ID
90493
FDA classified
2012-01-11
Recall event number
55294

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗