Open original FDA record ↗Recall record
- FDA identifier
- Z-0308-2027
- Company in FDA record
- Pentax of America Inc
- FDA classification date
- Oct 28, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Warnings added to the IFU for ultrasound endoscopes and water bottle assembly to ensure there is deaerated sterile water in the assembly prior to and during the procedure.
Source facts
- Action
- An Urgent Field Safety Notice dated 09/02/2026 was mailed to consignees. The notice informs consignees of two updates to the IFU of PENTAX Medical products, specifically the ultrasound endoscope devices and water bottle assembly. Consignees are to inform their staff of this notice and include the warning and updates to their IFUs. Additionally, consignees are instructed to change water for each procedure and fill the bottle approximately 2/3 full, or to the 200 marking on the bottle scale, with sterile water. If the water/balloon feeding function does not work consignees are to fill the water bottle with deaerated sterile water. PENTAX Medical can be contacted by phone at 1-800-431-5880 Monday through Friday from 8:30 AM EST to 5:00 PM EST, or by email at customeradvisories@pentaxmedical.com.
- Initiated
- 2026-09-09
- Root cause
- Unknown/Undetermined by firm
- Distribution
- Worldwide - US Nationwide and International distribution - Pending.
- Recall number
- Z-0308-2027
- Recall status
- Open, Classified
- CDRH record ID
- 223089
- FDA classified
- 2026-10-28
- Recall event number
- 99782