Open original FDA record ↗Recall record
- FDA identifier
- Z-0309-2018
- Company in FDA record
- Beckman Coulter Inc.
- FDA record creation date
- Jan 3, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
Potential for minor skin burn if the tool contacts the reagent storage interface board (B79917) when performing a pipettor alignment.
Source facts
- Action
- An Urgent Medical Device Recall letter, dated November 6, 2017, were sent to the affected customers via mail. The letter identifies the affected device, the reason for recall, impact, and actions to be taken. The firm states that customers may continue to operate their system normally. Customers are instructed to contact Technical Support if a device on the Access 2 instrument needs alignment. The alignment tool is only to be used by the Beckman Coulter service personnel and Beckman Coulter trained biomedical engineers. A Beckman Coulter service rep will contact customers to schedule a service visit. Customers are to share this information with their staff and retain a copy of the letter. Customers are to respond via one of the ways identified. Beckman Coulter should be contacted with any questions regarding the notice.
- Initiated
- 2017-11-06
- Root cause
- Device Design
- Terminated
- 2019-11-04
- Distribution
- Worldwide Distribution -- US, Argentina, Canada, Chile, China, Columbia, Costa Rica, France, Georgia, Germany, Ghana, Greece, Hong Kong, India, Italy, Japan, Kazakhstan, Mexico, Morocco, Myanmar, Nepal, Philippines, Poland, Russia, Rwanda, Saudi Arabia, Slovakia, South Africa, South Korea, Thailand, Turkey, United Kingdom, and Vietnam.
- Recall number
- Z-0309-2018
- Recall status
- Terminated
- CDRH record ID
- 159895
- FDA record created
- 2018-01-03
- Recall event number
- 78613